37-Kilodalton/83-kilodalton RNase L isoform ratio in peripheral blood mononuclear cells: analytical performance and relevance for chronic fatigue syndrome

Comment on: RNase L levels in peripheral blood mononuclear cells: 37-kilodalton/83-kilodalton isoform ratio is a potential test for chronic fatigue syndrome.[Clin Diagn Lab Immunol. 2003]

 

A French group has reported results (5) supporting the use of the RNase L 37-kDa/83-kDa ratio (37/83 R) in peripheral blood mononuclear cells (PBMC) as a diagnostic test for chronic fatigue syndrome (CFS). More recently (6), the same group cautioned about the diagnostic value of the 37/83 R, based on a small patient follow-up study which was likely to indicate analytical variability among duplicate assays, lack of reproducibility over time, and a weak correlation with the multidimensional fatigue inventory (MFI) score. Because of our long-term experience with this assay, we would like to offer some comments.

First, we tested the analytical performance of the 37/83 R assay according to CLSI (formerly NCCLS) procedure EP5-A (4), with control samples at three different levels made of extracts of the monocytic U937 cell line spiked with various concentrations of recombinant RNase L. The guideline protocol involves assaying the samples in duplicate twice daily over a total period of 20 days. The results summarized in Table Table11 indicate that both within- and between-run variation does not exceed 13%. In another series of experiments, we assayed eight patient samples in duplicate (average 37/83 R ranging from 0.5 to 245). Although in accordance with the NCCLS protocol results, the variation did not exceed 12% for samples with 37/83 R levels up to 20, and it rose significantly to 30% and more for samples with 37/83 R levels above 20. This should be expected, because beyond this level, more than 70% of the 83-kDa isoform is cleaved, and consequently, the faint 83-kDa band is difficult to scan with accuracy. Thus, in our opinion, the lower level of correlation between the duplicate assay results observed with the CFS group versus those with the controls (6) reflects the prevalence of high 37/83 R levels in the CFS group rather than a low test reproducibility as claimed by these authors. This is further supported by the good correlation found for the control group (r = 0.95). During validation, the lowest detectable ratio measured with a sample containing the 83-kDa isoform only was estimated (3 independent experiments with 26 replicates each) to be 0.13 ± 0.06 (average ± three standard deviations). Thus, the clinical cutoff ratio of 0.4 found by the authors (5, 6) to best discriminate CFS patients from controls falls within the measurable range.

You can read the rest of this comment here: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1247842/

 

Source: Frémont M, Vaeyens F, Herst CV, De Meirleir K, Englebienne P. 37-Kilodalton/83-kilodalton RNase L isoform ratio in peripheral blood mononuclear cells: analytical performance and relevance for chronic fatigue syndrome. Clin Diagn Lab Immunol. 2005 Oct;12(10):1259-60; author reply 1260. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1247842/ (Full article)

 